Beckman Coulter Optima L-100XPultracentrifuge
The Beckman Coulter Optima L-100 XP Ultracentrifuge brings you fast turn arounds
thanks to its powerful onboard software, efficient methods, and innovative rotors.
This ultra-centrifuge features an intuitive touch screen to help users be more effective,
almost a limitless number of user programs and will even simulate results
before performing separations to help check optimization.
The Beckman Coulter Optima L-100 XP Ultracentrifuge features an energy efficient air cooled drive,
constructed of high performance materials that allow it to reach speeds of up to 100,000 rpm,
and a maximum g-force of 802400 x g.
The drive of this ultracentrifuge is also imbalance tolerant, and can safely accommodate tubes
that are under or over filled by as much as 10%.
The Beckman Coulter Optima L-100 XP Ultracentrifuge has 10 acceleration and 11 deceleration profiles,
that include 9 slow rates as well as coasting to a stop without brake, or full dynamic braking from set speed.
This floor model centrifuge has quiet operation at <53 dBa,
so you can have it in the lab nearby without being too distracting.
The temperature range of this ultracentrifuge can be set between 10 to 40°C in
increments of 1° with a thermo electric refrigeration system and a moisture purging vacuum system.
Contact us today to learn more about our Beckman Ultracentrifuge.
Beckman Coulter L-100XP Ultracentrifuge Specifications:
Approximate Acceleration Time: 10 acceleration profiles–9 slow rates or maximum acceleration from 0 to set speed
Approximate Deceleration Time: 11 deceleration profiles–9 slow rates, coasting to a stop without brake, or full dynamic braking from set speed
g Force: 802,400 g (Type 100 Ti)
Imbalance Tolerance: ±5 ml or 10% whichever is greater
Max Speed: 100,000 rpm
Maximum Heat Dissipation into Room under Steady-State Conditions: 1.0kW/hr (3400 BTU/hr)
Noise Level 1 Meter in Front of Centrifuge: <53 dBa
Set Speed: 0 to 40° C in 1° increments
Temperature Control: ±0.5° of set temperature (after equilibration)
Ambient Temperature Range: 10 to 40°C ambient
Clearances Required: 5.1 cm (2.0 in) sides; 15.2 cm (6.0 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Time – Actual Display: Indicates time remaining in timed runs, time elapsed in Hold runs, or estimated time remaining in Ω2t runs
Vacuum: Moisture-purging vacuum system
Finish: Control panel: coated polycarbonate. Top surface: urethane paint. Other surfaces: general purpose paint.
Dimensions H x W x D 49.5 x 37 x 26.5 in (125.7 x 94 x 67.3 cm)
Weight: 465 kg (1025 lb)